PBS-204

Pulverised Iron Ore, 1kg jar

PBS-204 is a pulverised iron ore certified reference material (CRM), sourced from the Pilbara region of Western Australia and certified in accordance with ISO Guide 35 and ISO 17034. This CRM is specifically designed for quality control (QC) of XRF and TGA analysis, with precise characterization of major and trace elements in iron ore. 

 Key Features 

  • Certified concentrations for major elements (Fe, SiO₂, Al₂O₃) and trace components (Mn, P, S, TiO₂).
  • High iron content (65.75% Fe) with low gangue minerals, ideal for iron ore analysis. 
  • Produced via multi-stage crushing, pulverisation, and homogenisation to ensure uniformity. 
  • Supplied in 1kg press-sealed plastic jars, with 10g sachets available upon request for blind laboratory submissions. 
  • Stable under normal storage conditions when kept sealed. 

 Key Certified Values 

Analyte Certified Value (y) Standard Deviation (s) 95% Confidence Interval (CI)
1 SD 1 SD Within Lab Lower Upper
Fe 65.75 0.163 0.134 65.69 65.88
SiO2 1.14 0.040 0.010 1.10 1.16
Al2O3 0.98 0.022 0.012 0.96 0.99
TiO2 0.060 0.0051 0.0019 0.055 0.062
Mn 0.019 0.0008 0.0005 0.019 0.020
P 0.124 0.0011 0.0007 0.123 0.125
S 0.014 0.0008 0.0003 0.013 0.015
MgO 0.040 0.0098 0.0037 0.033 0.047
K2O 0.003 0.0011 0.0008 0.002 0.004
Na2O 0.010 0.0108 0.0034 0.002 0.029
LOI371 3.00 0.029 0.018 2.99 3.04
LOI425 3.09 0.014 0.014 3.08 3.10
LOI650 3.30 0.022 0.022 3.29 3.31
LOI1000 3.46 0.026 0.022 3.44 3.47
LOI371 to 425 0.08 0.018 0.007 0.07 0.10
LOI425 to 650 0.20 0.024 0.009 0.18 0.23
LOI650 to 1000 0.16 0.015 0.012 0.15 0.17

 Intended Use 

  • QC monitoring for XRF/TGA analysis of iron ore. 
  • Method validation for laboratories processing standard iron ores. 
  • Calibration checks to ensure analytical accuracy and precision. 

Pulverised Iron Ore

SKU: 10204

This product is out of stock.

Alternate Packing Options

CONTACT US
Share