PBS-310

Pulverised sediments and low grade BIF, 1kg jar

PBS-310 is a certified reference material (CRM) prepared from pulverised sediments and low-grade banded iron formation (BIF) sourced from the Pilbara region of Western Australia. Spiked with elevated levels of NaCl and Ca, this CRM is designed for laboratories performing XRF fusion and TGA (LOI) analysis. It was prepared and certified in accordance with ISO 17034:2016 and ISO Guide 35:2017, ensuring reliable performance in quality control and validation applications.

Key Features

  • Certified for major and minor oxides in sediment-hosted BIF material.
  • Produced by multi-stage drying, pulverising, and homogenisation to ensure excellent uniformity.
  • Certified through an inter-laboratory study involving multiple ISO 17025-accredited laboratories.
  • Supplied in 1kg press-sealed plastic jars, with 10g sachets available upon request for blind laboratory submissions. 
  • Stable under ambient storage conditions with a 10-year certificate validity.

Key Certified Values

Analyte

Certified Value (y)

Standard Deviation

95% Confidence Interval (CI)

1 SD (s)

1 SD Within Lab (sw)

lower

upper

Fe

22.98

0.161

0.078

22.85

23.11

SiO2

30.24

0.139

0.064

30.14

30.34

Al2O3

14.20

0.091

0.031

14.13

14.27

TiO2

0.830

0.0080

0.0056

0.825

0.835

Mn

1.73

0.038

0.011

1.70

1.76

CaO

2.84

0.026

0.011

2.82

2.86

P

0.040

0.0008

0.0007

0.040

0.040

S

0.027

0.0017

0.0006

0.026

0.028

MgO

2.64

0.033

0.011

2.62

2.66

K2O

0.981

0.0081

0.0034

0.975

0.987

Na2O

1.70

0.040

0.016

1.67

1.73

Cl

1.73

0.187

0.074

1.58

1.88

LOI425

3.44

0.467

0.068

2.81

4.07

LOI650

10.04

0.155

0.052

9.91

10.17

LOI1000

10.93

0.097

0.052

10.85

11.01

LOI425 to 650

6.57

0.446

0.054

5.97

7.17

LOI650 to 1000

0.93

0.187

0.055

0.77

1.09

Intended Use

  • Monitoring and verifying accuracy and precision of XRF fusion and TGA (LOI) workflows.
  • Supporting internal quality control programs in geochemistry and mining labs.
  • Suitable for calibration and bias detection in sediment-rich and low-grade Fe-bearing matrices.

Pulverised Iron Ore

SKU: 10310

Add to Cart

$403.60

Alternate Packing Options

CONTACT US
Share